PS300 Series - High Voltage Power Supply
|
|
PS350
|
The PS300 High Voltage Power Supply series consists of three efficient, microprocessor-controlled, switching DC power supplies capable of delivering 25 W of output power at voltages up to 5 kV. All models have 0.001 % regulation and 0.05 % accuracy, and the output voltage can be adjusted with 1 V resolution over the entire operating range. |
| All PS300 series supplies have rear-panel SHV connectors. Optional cables allow connection with standard high-voltage connectors (SHV or MHV). A three-position, high-voltage enable switch on the front panel prevents the high voltage from being turned on under computer control unless the switch has been manually armed. A highly visible red LED always indicates when the high voltage is on. Convenient front-panel LED displays indicate both the voltage and current delivered to the load at all times. |
PS350
|
Limits and Trips
The PS300 high voltage power supplies have an adjustable voltage limit which prevents the high voltage from being inadvertently set above a safe level. An independent current limit lowers the voltage setting until the current drawn by the load does not exceed the limit setting. Additionally, a current trip may be set which turns off the high voltage when the current limit is exceeded. After a trip, the unit can be configured to either attempt to turn the high voltage back on, or to leave it off until a manual reset.
Adjustable Polarity
The output polarity of the PS300 High Voltage Power Supply series is set with a rear-panel switch. The polarity cannot be controlled via the computer interface and can only be changed while the unit is off. Output polarity is always displayed on the front panel with the voltage level.
PS300 Block Diagram
Voltage and Current Monitors
Two rear-panel BNC connectors provide voltage and current monitoring capabilities. A 0 V to 10 V output corresponds to 0 % to 100 % of full scale. These outputs are capable of driving 10 mA and have a 1 Ω output impedance.
The voltage monitor output can also be configured as a voltage control input. In this mode, a 0 V to 10 V signal applied at the input will cause the output high voltage to vary between 0 % and 100 % of full scale. The output is updated at a rate of 16 Hz. Additionally, the bandwidth of the voltage control input is limited by the overall slew rate specification of the instrument (0 % to 100 % of full scale in less than 0.3 s under full load).
Stored Settings
All instrument settings are stored in non-volatile memory so that the instrument will remember its previous state when powered up. For safety, the high voltage is always initially turned off. Up to nine complete instrument configurations, including limit and trip settings, can be stored in memory.
Optional GPIB Interface
An optional GPIB interface allows control of all instrument functions except output polarity. The required common commands of the IEEE-488.2 standard are supported. The PS300 High Voltage Power Supply series can be configured to generate service requests (SRQ) in the event of limit or trip conditions.
PS300 Series Specifications
| |
||
| Model | Output Voltage(±) | Max.Current |
| PS310 | 0 to ±1.25 kV | 20 mA |
| PS325 | 0 to ±2.5 kV | 10 mA |
| PS350 | 0 to ±5 kV | 5 mA |
| PS355 | 0 to -10 kV | 1 mA |
| PS365 | 0 to +10 kV | 1 mA |
| PS370 | 0 to -20 kV | 0.5 mA |
| PS375 | 0 to +20 kV | 0.5 mA |
| |
|
| Voltage Output | |
| Voltage set accuracy | 0.01 % + 0.05 % of full scale |
| Voltage display accuracy | Vset accuracy ±1 V, typ. (±2 V, max.) |
| Voltage resolution | 1 V (set and display) |
| Voltage resettability | 1 V |
| Voltage limit range | 0 to 100 % of full scale |
| Voltage regulation | 0.001 % for ±10 % line change |
| 0.005 % for 100 % load change | |
| Specifications apply for >0.5 % (full load) to >1 % (no load) of full scale voltage. | |
| Output ripple (Vrms) | 0.0015 % of full scale (typ.), 0.002 % of full scale (max.) |
| Current limit range | 0 to 105 % of full scale |
| Trip current | 10 µA (min.) |
| Trip response time | <10 µs |
| Current set accuracy | 0.01 % + 0.05 % of full scale |
| Current resolution | 10 µA (PS310 and PS325) |
| 1 µA (PS350)) | |
| Current display accuracy | ±10 µA (typ.), ±20 µA (max.) (PS310 and PS325) |
| ±1 µA (typ.), ±2 µA (max.) (PS350) | |
| Stability | 0.01 % per hr., <0.03 % per 8 hrs. |
| Temperature drift | 50 ppm/°C, 0 °C to 50 °C (typ.) |
| Protection | Arc and short circuit protected. Programmable voltage limit, current limit and current trip |
| Recovery time | 12 ms for 40 % step change in load current (typ.) |
| Discharge time | <6 s (to <1 % of full-scale voltage with no load, typ.) |
| |
|
| Monitor Output | |
| Output scale | 0 to +10 V for 0 to full-scale output regardless of polarity |
| Current rating | 10 mA (max.) |
| Output impedance | <1 Ω |
| Accuracy | 0.2 % of full scale |
| Update rate | 8 Hz |
| |
|
| External Voltage Set | |
| Input scale | 0 to +10 Volts for 0 to full-scale output regardless of polarity |
| Input impedance | 1 MΩ |
| Accuracy | 0.2 % of full-scale |
| Update rate | 16 Hz |
| Output slew rate | <0.3 s for 0 to full-scale under full load |
| |
|
| Mechanical Specifications | |
| HVconnector | SHV male (Kings Type 1704-1) |
| Mating connector | SHV female (Kings Type 1705-14) |
| Dimensions | 8.1" × 3.5" × 16.0" (WHL) |
| Weight | 8 lbs. |
| Power | 50 W, 100/120/220/240 VAC, 50/60 Hz |
| Warranty | One year parts and labor on defects in materials and workmanship |