SR-800 Extended area Blackbody
|
|
Features
The SR-800 sets a new standard for accuracy and uniformity in extended area blackbodies. Temperature measurement and calibration are both calculated in the radiation head itself and then transferred digitally to the controller. The result is a more accurate, stable and flexible system that gives reliable NIST-traceable results. The SR-800 operates as a complete standalone system or may be integrated into a larger test system.
- Interchangeable heads with the same controller
- Ability to control up to four motorized devices
- Calibrated radiation heads independent of the controller for easy maintenance
- Customized head dimensions upon request
- Ethernet, GPIB and RS232 communication ports
- Compact and portable controller
- Remote control and maintenance over a local area network
- 19” rack mount kit included
- Large color LCD display
- Windows CE operating system
- Touch screen user interface
Certification
- NIST-traceable calibration
- Certified to MIL-T-28800D, CE, and FCC
Specifications
| SR-800 Model | 2D/A | 4D/A | 7D/A | 12D/A | 14A | 20A |
| Aperture | 2" dia. | 4"x4" | 7"x7" | 12"x12” | 14"x14” | 20"x20” |
| Absolute temp. range | 0ºc to 125ºc | 0ºc to 125ºc | 5ºc to 90ºc | 10ºc to 80ºc | 10ºc to 80ºc | 10ºc to 80ºc |
| Differential temp. range | -25ºc to 100ºc | -25ºc to 100ºc | -20ºc to 65ºc | -15ºC to 55ºc | -------- | |
| Set point & readout resolution | 0.001ºc | |||||
| Uniformity2 | ±0.01ºc | ±0.01ºc | ±0.015ºc | ±0.03ºc | ±0.02ºc | ±0.02ºc |
| Absolute Temp. Accuracy3 | 0.008ºc @ Tabsolute<50ºc
0.015ºc @ Tabsolute>50ºc |
0.008ºc @ Tabsolute<50ºc
0.03ºc @ Tabsolute>50ºc |
||||
| Differential Temp. Accuracy3 | 0.008ºc @ ΔT<±22ºc
0.015ºc @ΔT>22ºc |
-------- | ||||
| Total system uncertainty | 0.0025 x (25ºc-T) for T<15ºc
0.025ºc for 15ºc<T<35ºc 0.0015x(T-25ºc)+0.01ºc for 35ºc<T |
|||||
| Stability | ±0.002ºc @ ΔT <±10ºc | ±0.005ºc @ ΔT <±10ºc | ||||
| Emissivity | 0.97 ±0.02 | |||||
| Settling time4 | 30 sec | |||||
| Slew rate | 0.5ºc/sec | 0.33ºc/sec | 0.22ºc/sec | 0.17ºc/sec | 0.1ºc/sec | 0.1ºc/sec |
| Head size (mm) | 126x
133x 80 |
161x
190x 201 |
224x
225x 265 |
360x
347x 380 |
410x
170x 480 |
570x
170x 600 |
| Head weight | 1kg | 5kg | 10kg | 28kg | 36kg | 65kg |
| Power consumption | 200w | 200w | 250w | 1200w | 2400w | 4500w |
Options
| SR-800 Model | 2D/A | 4D/A | 7D/A | 12D/A | 14A | 20A |
| ET – Extended Temp. | Absolute temp. range | 0ºc to 175ºc | 5ºc to 125ºc | 5ºc to 150ºc | ||
| Differential temp. range | -25ºc to 150ºc | -20ºc to 100ºc | ||||
| HU – High uniformity | ||||||
| Uniformity2 | ±0.008ºc | ±0.014ºc | ||||
| LT – Low Temp | ||||||
| Absolute temp. range | -40ºc to 150ºc | -40ºc to 150ºc | ||||
| Differential temp. range | -65ºc to 125ºc | -65ºc to 125ºc | ||||
| WTR – Wide Temp. Range5 | ||||||
| Absolute temp. range | -20ºc to 175ºc | |||||
| Differential temp. range | -45ºc to 150ºc | |||||
| HE – High Emissivity | ||||||
| Emissivity | 0.99 ±0.01 | |||||
| FS – Fast Slew Rate | ||||||
| Slew rate | 0.33ºc/sec | 0.25ºc/sec | ||||
| CH Environmental chamber | ||||||
| Chamber temp. | -30ºc to 70ºc | |||||
| Absolute temp. range | -40ºc to 80ºc | |||||
| Differential temp. range | -10ºc to 40ºc | |||||
| Accuracy3 | 0.01ºc x ?T or 0.05ºc whichever is greater | |||||