Home Gas Sampling & Analysis Systems SRS PPR100* Vacuum Process Monitoring System

SRS PPR100* Vacuum Process Monitoring System

Product Code:
15064
Manufacturer: Stanford Research Systems
  • 0.01, 0.1, 1.0 or 10 Torr inlet pressure 
  • 3 × 10-5 Torr-L/s flow rate 
  • 2 second response time (0.1 Torr) 
  • 100, 200 or 300 amu systems 
  • Field-replaceable electron multiplier and filament 
  • 6 orders of magnitude dynamic range 
  • RGA Windows software

More info and finance questions

Request a Quote

Please type the letters and numbers below
 The Stanford Research Systems PPR Process Monitoring System is designed for inline process monitoring and diagnosis. Two paths are provided to the residual gas analyser (RGA): a high conductivity path for monitoring base vacuum, and a pressure reducing path for monitoring the process at operating pressure.

The pressure reducing path contains a micro-hole orifice which is designed to operate at one of the following pressures: 0.01, 0.1, 1 or 10 Torr. It reduces the sample pressure to the operating pressure of the RGA (about 10-6 Torr). This pressure drop is maintained by the pumping system which consists of a hybrid turbomolecular pump and a diaphragm pump. Both pumps are oil free and will not contaminate your process.

The inlet assembly that attaches to your process chamber can be configured in several ways (see drawings). The system includes an RGA, by-pass valve assembly and Tee, a controller, turbo pump, diaphragm pump, and Windows software program for data acquisition and control.

The software is used to operate the instrument in various modes, including analog scan, histogram mode, and pressure vs. time mode. Both Faraday cup and electron multiplier detectors are standard with the PPR system. The electron multiplier provides additional sensitivity and higher scan speeds. The PPR system is shipped completely assembled and calibrated, and is ready to attach to your vacuum process chamber.

Performance  
Gas flow 3 × 10-5 mbar-L/s with pressure reduction inlet active
Response time
 
2 seconds at 0.1 mbar inlet pressure (scales linearly with pressure)
Start-up time 8 minutes nominal
   
Connections  
Inlet 2.75" CF flange, rotatable with through holes
Turbo pump to controller 6 ft. cable (provided) 
 
Turbo to backing pump 6 ft. flexible hose (provided), 1/4" ID × 7/16" OD 
 
Computer interface RS-232C (28,800 baud, 9-pin D connector)
Software RGA Windows application
   
Pumps  
High vacuum 
 
Hybrid turbomolecular/drag pump, 70 L/s, ultimate pressure 2 × 10-9 mbar
Backing 
 
Diaphragm pump with ultimate pressure less than 1 mbar. Protection class IP44
Cooling Requires forced air cooling
   
General  
Power requirements 110 VAC @ 60 Hz, 220 VAC @ 50 Hz (not field selectable), 300 W 
 
Dimensions Vary with configuration (see sample configurations)
Weight 
 
16 lbs. (turbo pump, by-pass valve and Tee, RGA) 
33 lbs. (diaphragm pump and controller)
Warranty 
 
One year parts and labor on defects in material or workmanship. Pump seals and diaphragm warranted for 90 days.
Write Your Own Review
You're reviewing:SRS PPR100* Vacuum Process Monitoring System
Your Rating
Photo Product SKU
SRS BGA244 Binary Gas Analyser SRS BGA244 Binary Gas Analyser - >
SRS UGA100* Universal Gas Analyser SRS UGA100* Universal Gas Analyser - >
SRS UGA200* Universal Gas Analyser SRS UGA200* Universal Gas Analyser - >
SRS UGA300* Universal Gas Analyser SRS UGA300* Universal Gas Analyser - >
SRS QMS100* High Pressure Sampling Systems SRS QMS100* High Pressure Sampling Systems - >
SRS QMS200* High Pressure Sampling Systems SRS QMS200* High Pressure Sampling Systems - >
SRS QMS300* High Pressure Sampling Systems SRS QMS300* High Pressure Sampling Systems - >
SRS PPR200* Vacuum Process Monitoring System SRS PPR200* Vacuum Process Monitoring System - >
SRS PPR300* Vacuum Process Monitoring System SRS PPR300* Vacuum Process Monitoring System - >