Tempo T27603 MM Optical Loss Test Set + VisiTester
£2,381.00 £2,857.20
An Optical Loss Test Set used to test power, insertion loss, continuity and faults on multimode fibre optic systems at 850/1300 nm. In a small package, it provides a powerful solution to improve test flexibility and productivity.
The 650 nm VFL Visitester feature mixes a VFL in with the test signal during Autotest. In addition to operating as a typical VFL, it gives positive visual identification of a fiber to be connected for loss testing.
- Easy to use, slim, rugged & versatile
- EF compliant multimode sources
- Excellent source stability
- Large clear LCD is sunlight readable & backlit
- Large memory with Text Tag & time stamp
- One button memory dump to USB key
- KITS™ reporting software for acquisition and reporting
- Autotest power meter
- 1 % accuracy & 31 calibrated λ, ISO/IEC 17025 traceable
- Interchangeable connectors with dust cap & tilt bail
- Multi-Fiber ID Test tone generator / detector
- Long alkaline battery life with micro USB external power / charger
Attribute Name | Attribute Value |
---|---|
Detector type (meter) | InGaAs |
Response wavelength (meter) | 600 ~ 1700 nm |
Calibration wavelengths (meter) | 780, 820, 850, 980, 1270, 1290, 1300, 1310, 1330, 1350, 1370, 1390, 1410, 1430, 1450, 1470, 1490, 1510, 1530, 1550, 1570, 1590, 1610, 1625, 1650 nm |
Power range (meter) | +10 ~ -60 dBm (780 ~ 980 nm) / +10 ~ -70 dBm (1270 ~ 1650 nm) |
Mid-range linearity (meter) | 0.04 dB (typical) |
Total uncertainty (meter) | 0.3dB (max) |
ID tone detection & Autotest sensitivity (meter) | Up to -50 dBm |
Modulated output (source) | 270 Hz, 1 kHz, 2 kHz ± 2 %, 12 Multi-Fiber ID tones |
Output wavelength (source) | Port 3: 850-300 nm LED, VisiTester |
Output power (source) | -23 dBm ±1 dB (@ 62.5/125 µm for 850, 1300 nm) |
Adjust output power (source) | Fixed (for 850, 1300 nm) / Adjustable over 7 dB in 0.01 dB steps (for VisiTester) |
Short term stability for 15 min, typical ±2 °C, after warm up, ORL < -25 dB (source) | 0.01 dB (for 850, 1300 nm) |
Stability over temperature (source) | 0.35 dB (typical for 850, 1300 nm) |
Wavelength drift (source) | 0.4 nm per °C (typical for 850, 1300 nm) |
Mode controlled (source) | 850, 1300 nm (50/125µm compliant: IEC 61280-4-1 {Ed.1.0}, TIA 526-14A & TIA TSB-178) |
Reconnection repeatability @ 95% confidence (source) | 0.05 dB (for 850, 1300 nm), 0.1 dB (for VisiTester) |
VisiTester wavelength | 650 ± 5 nm |
VisiTester spectrum width @FWHM | 3 nm |
VisiTester wavelength drift | 0.1 nm per °C |
VisiTester power | +2 dBm ± 1 dB @ MMF |
VisiTester laser safety | Class 1, IEC60825-2 |
VisiTester blink rate (manual) | CW or 2 Hz |
Accuracy (meter) | 1 % (0.06 dB) |
Calibration validity (ISO/IEC 17025) | 3 years |
Resolution (meter) | 0.01 dB |
Polarization sensitivity (meter) | < 0.05 dB (typical) |
Autotest | 1-way: compatible with Kingfisher's 1-way Loss Test Sets & Power Meters (except T9600 series) |
Tone detection (meter) | 150 ~ 9900 Hz ± 1 % |
Battery life | Laser/LED source: 90/80 hours (typical) in Autotest |
Power | Battery / External power via USB |
Battery type | 2x AA (LR6) Alkaline / 2x AA NiMH AA |
USB interfaces | USB-micro for power & software / USB-A for memory key |
Auto off | Selectable |
Software | KITS (free download) |
Memory | 1,000 four λ tests with date & time / Unlimited on USB |
Case | Polycarbonate (body) / rubber (edges & corners), moisture resistance, 1-meter drop tested |
Display size | 74 mm x 55 mm |
Size | 190 mm x 105 mm x 35 mm |
Unit weight | 0.42 kg |
Shipping weight | 1.5 Kg |
Operating / storage temperature | -15 to 55 °C / -25 to 70 °C |
Relative humidity | 0 ~ 95 % |
Warranty | 3 years |